Electronic Pipettes, the Acura® electro Line of 926 Micropipettes, 936 Macropipettes and 956 Multichannel Pipettes
Overview
The Accura® electro Line of Micropipettes, Macropipettes and Multichannel Pipettes is the latest generation of new technology making electronic pipetting simpler and safer than ever.
The microprocessor controlled instrument includes software that avoids any complicated programming steps, so it is easy to get quick access to six working modes and many volumes to select from.
All the Accura® electro Micropipettes, Macropipettes and Multichannel Pipettes share these features:
Optimal ergonomics, light weight, perfect hand fitting
Includes microprocessor-controlled motor and rechargeable battery pack.
Easy, intuitive, self-teaching program
Large display, reversible right / left reading
Quick change battery pack
Fast charge battery, long charge life
Adjustable patented tip ejector fitting most tips
Pipetting cycle counter
Interchangeable volumetric modules available as accessories
Volumes ranging from 0.1 µL to 10 mL.
Outstanding performance, accuracy, precision and stability
Two year warranty
Features of the Acura® electro 926 micropipettes:
Eight models cover the volume range 0.1 to 1000 µL
Reduced shaft length and conical end improve driveability
Easy access to microtubes and microplates
Optional longer volumetric modules
Simple conversion by user at any time
Features of the Acura® electro 936 macropipettes:
Three models cover the volume range 0.1 to 10 mL
Interchangeable nozzle protection filter
Adapters for Pasteur pipette and straw tips
Features of the Acura® electro 956 multichannel pipettes:
Large selection of 8- and 12-channel models
Cover the volume range 0.5 to 350 µL.
Sequential tip ejection
Also fit any single channel volumetric module
Because of the outstanding shape and balance, the Accura® electro Line of Micropipettes, Macropipettes and Multichannel Pipettes handle very much like manual instruments, but comfort is so much improved that productivity and result consistency are simply outstanding.
It is easy to cycle through six modes: Forward (for all applications), Reverse (especially for viscous or foaming liquids), Stepper (for aliquot samples), Dilution, Tactile and Mixing.
One control unit fits a wide choice of volumetric assemblies available separately, extending the working possibilities and making electronic pipetting affordable. Disassembly and re-assembly is done without any tools, for maximum flexibility. Calibration specifics are retained in instrument memory. Maintenance is kept to a minimum. If sterilization is needed, volumetric modules can be autoclaved at 121 °C.
The NiMH battery pack is quickly exchangeable, charges fast, lasts long, and goes into automatic energy saving standby when it is left unused.
The shaft height adjustment system called Justip™, controlled by click stops, allows a wide selection of tips to tightly fit the nozzle. Tip ejection is easier than ever.
Each instrument has its own serial number and passes strict performance requirements as attested to by an individual Quality Control Certificate that is supplied.
Specifications (Click here):
Electronic Pipettes, the Acura® electro Line of 926 Micropipettes, 936 Macropipettes and 956 Multichannel Pipettes
Specifications
Micropipette Model:
Acura® 926.0002
Acura® 926.0010
Acura® 926.001Y
Acura® 926.0020
Acura® 926.0050
Acura® 926.0100
Acura® 926.0200
Acura® 926.1000
Volume: (µL)
0.1 - 2
0.5 - 10
0.5 - 10
1 - 20
2.5 - 50
5 - 100
10 - 200
50 - 1000
Division: (µL)
0.01
0.05
0.05
0.1
0.25
0.5
1.0
5.0
Inaccuracy (relative error) Minimum volume: (%)
< ± 2.5 at 0.5 µL
< ± 1.2 at 10% of nominal volume
< ± 1.2 at 10% of nominal volume
< ± 1.2 at 10% of nominal volume
< ± 1.0 at 10% of nominal volume
< ± 1.0 at 10% of nominal volume
< ± 1.0 at 10% of nominal volume
< ± 0.8 at 10% of nominal volume
Inaccuracy (relative error) Mid volume: (%)
< ± 1.2
< ± 0.8
< ± 0.8
< ± 0.6
< ± 0.6
< ± 0.6
< ± 0.6
< ± 0.5
Inaccuracy (relative error) Maximum volume: (%)
< ± 0.9
< ± 0.6
< ± 0.6
< ± 0.5
< ± 0.5
< ± 0.5
< ± 0.4
< ± 0.4
Imprecision (CV) Minimum volume: (%)
< 2.5 at 0.5 µL
< 1.5 at 10% of nominal volume
< 1.7 at 10% of nominal volume
< 1.2 at 10% of nominal volume
< 0.7 at 10% of nominal volume
< 0.7 at 10% of nominal volume
< 0.6 at 10% of nominal volume
< 0.4 at 10% of nominal volume
Imprecision (CV) Mid volume: (%)
< 1,5
< 0.7
< 0.8
< 0.4
< 0.3
< 0.3
< 0.2
< 0.15
Imprecision (CV) Maximum volume: (%)
< 0.8
< 0.35
< 0.4
< 0.3
< 0.25
< 0.2
< 0.15
< 0.1
Tip style: (µL)
ultra 10
ultra 10
200
200
200
200
200
1000
Specifications
Macropipette and 8-Channel or 12-Chanel Model:
Acura® 936.02 Macro-pipette
Acura® 926.05 Macro-pipette
Acura® 926.10 Macro-pipette
Acura® 956.08.010 956.12.010 8 or 12 Multi-channel
Acura® 956.08.050 956.12.050 8 or 12 Multi-channel
Acura® 956.08.200 956.12.200 8 or 12 Multi-channel
Acura® 956.08.300 956.12.300 8 or 12 Multi-channel
Volume: Macro (mL) Multi (µL)
0.1 - 2
0.25 - 5
0.5 - 10
0.5 - 10
2.5 - 50
10 - 200
20 - 350
Div: Macro (mL) Multi (µL)
0.01
0.05
0.05
0.05
0.25
1.0
5.0
Inaccuracy (relative error) Minimum volume: (%)
< ± 1.5 at10% of nominal volume
< ± 1.2 at10% of nominal volume
< ± 1.0 at10% of nominal volume
< ± 3.5 at10% of nominal volume
< ± 1.0 at10% of nominal volume
< ± 0.9 at10% of nominal volume
< ± 1.0 at10% of nominal volume
Inaccuracy (relative error) Mid volume: (%)
< ± 1.0
< ± 0.8
< ± 0.5
< ± 1.5
< ± 0.9
< ± 0.7
< ± 0.8
Inaccuracy (relative error) Maximum volume: (%)
< ± 0.5
< ± 0.5
< ± 0.5
< ± 1.0
< ± 0.8
< ± 0.6
< ± 0.6
Imprecision (CV) Minimum volume: (%)
< 0.6 at 10% of nominal volume
< 0.6 at 10% of nominal volume
< 0.5 at 10% of nominal volume
< 3.0 at 10% of nominal volume
< 1.0 at 10% of nominal volume
< 0.6 at 10% of nominal volume
< 0.6 at 10% of nominal volume
Imprecision (CV) Mid volume: (%)
< 0.3
< 0.3
< 0.2
< 0.9
< 0.6
< 0.4
< 0.4
Imprecision (CV) Maximum volume: (%)
< 0.15
< 0.15
< 0.15
< 0.7
< 0.4
< 0.25
< 0.25
Tip style: Macro (mL) Multi (µL)
2
5
10
utra 10
200
200
350
Number
Description
Price $
926.0002
Acura® electro 926 Micropipette, 0.1 – 2 µL, supplied with QC certificate, operating instructions, and Socorex Qualitips® pipette tip samples (W870902)
934.03
926.0002U
Acura® electro 926 Micropipette, 0.1 – 2 µL, Initial Package includes electronic pipette, individual QC certificate and operating instructions, pipette charging stand, fast charger with AC adapter and cord, additional battery pack, and Socorex Qualitips® pipette tip samples (W870902-A)
1,333.31
926.0010
Acura® electro 926 Micropipette, 0.5 – 10 µL, supplied with QC certificate, operating instructions, and Socorex Qualitips® pipette tip samples (W870904)
934.03
926.0010U
Acura® electro 926 Micropipette, 0.5 – 10 µL, Initial Package includes electronic pipette, individual QC certificate and operating instructions, pipette charging stand, fast charger with AC adapter and cord, additional battery pack, and Socorex Qualitips® pipette tip samples (W870904-A)
1,333.31
926.0010Y
Acura® electro 926 Micropipette, 0.5 – 10 µL for 200 µL tip style, supplied with QC certificate, operating instructions, and Socorex Qualitips® pipette tip samples (W870906)
934.03
926.0010YU
Acura® electro 926 Micropipette, 0.5 – 10 µL for 200 µL tip style, Initial Package includes electronic pipette, individual QC certificate and operating instructions, pipette charging stand, fast charger with AC adapter and cord, additional battery pack, and Socorex Qualitips® pipette tip samples (W870906-A)
1,333.31
926.0020
Acura® electro 926 Micropipette, 1 – 20 µL, supplied with QC certificate, operating instructions, and Socorex Qualitips® pipette tip samples (W870908)
934.03
926.0020U
Acura® electro 926 Micropipette, 1 – 20 µL, Initial Package includes electronic pipette, individual QC certificate and operating instructions, pipette charging stand, fast charger with AC adapter and cord, additional battery pack, and Socorex Qualitips® pipette tip samples (W870908-A)
1,333.31
926.0050
Acura® electro 926 Micropipette, 2.5 – 50 µL, supplied with QC certificate, operating instructions, and Socorex Qualitips® pipette tip samples (W870910)
934.03
926.0050U
Acura® electro 926 Micropipette, 2.5 – 50 µL, Initial Package includes electronic pipette, individual QC certificate and operating instructions, pipette charging stand, fast charger with AC adapter and cord, additional battery pack, and Socorex Qualitips® pipette tip samples (W870910-A)
1,333.31
926.0100
Acura® electro 926 Micropipette, 5 – 100 µL, supplied with QC certificate, operating instructions, and Socorex Qualitips® pipette tip samples (W870912)
934.03
926.0100U
Acura® electro 926 Micropipette, 5 – 100 µL, Initial Package includes electronic pipette, individual QC certificate and operating instructions, pipette charging stand, fast charger with AC adapter and cord, additional battery pack, and Socorex Qualitips® pipette tip samples (W870912-A)
1,333.31
926.0200
Acura® electro 926 Micropipette, 10 – 200 µL, supplied with QC certificate, operating instructions, and Socorex Qualitips® pipette tip samples (W870914)
934.03
926.0200U
Acura® electro 926 Micropipette, 10 – 200 µL, Initial Package includes electronic pipette, individual QC certificate and operating instructions, pipette charging stand, fast charger with AC adapter and cord, additional battery pack, and Socorex Qualitips® pipette tip samples (W870914-A)
1,333.31
926.1000
Acura® electro 926 Micropipette, 50– 1000 µL, supplied with QC certificate, operating instructions, and Socorex Qualitips® pipette tip samples (W870916)
934.03
926.1000U
Acura® electro 926 Micropipette, 50 – 1000 µL, Initial Package includes electronic pipette, individual QC certificate and operating instructions, pipette charging stand, fast charger with AC adapter and cord, additional battery pack, and Socorex Qualitips® pipette tip samples (W870916-A)
1,333.31
936.02
Acura® electro 936 Macropipette, 0.1 – 2 mL, supplied with QC certificate, operating instructions, and Socorex Qualitips® pipette tip samples (W870918)
978.59
936.02U
Acura® electro 936 Macropipette, 0.1 – 2 mL, Initial Package includes electronic pipette, individual QC certificate and operating instructions, pipette charging stand, fast charger with AC adapter and cord, additional battery pack, and Socorex Qualitips® pipette tip samples (W870918-A)
1,372.53
936.05
Acura® electro 936 Macropipette, 0.25 – 5 mL, supplied with QC certificate, operating instructions, and Socorex Qualitips® pipette tip samples (W870920)
978.59
936.05U
Acura® electro 936 Macropipette, 0.25 – 5 mL, Initial Package includes electronic pipette, individual QC certificate and operating instructions, pipette charging stand, fast charger with AC adapter and cord, additional battery pack, and Socorex Qualitips® pipette tip samples (W870920-A)
1,372.53
936,10
Acura® electro 936 Macropipette, 0.5 – 10 mL, supplied with QC certificate, operating instructions, and Socorex Qualitips® pipette tip samples (W870922)
978.59
936,10U
Acura® electro 936 Macropipette, 0.5 – 10 mL, Initial Package includes electronic pipette, individual QC certificate and operating instructions, pipette charging stand, fast charger with AC adapter and cord, additional battery pack, and Socorex Qualitips® pipette tip samples (W870922-A)
1,372.53
956.8.010
Acura® electro 956 8-Channel Micropipette, 0.5 – 10 µL, supplied with QC certificate, operating instructions, and Socorex Qualitips® pipette tip samples (W870924)
1,622.08
956.8.010U
Acura® electro 956 8-Channel Micropipette, 0.5 – 10 µL, Initial Package includes electronic pipette, individual QC certificate and operating instructions, pipette charging stand, fast charger with AC adapter and cord, additional battery pack, and Socorex Qualitips® pipette tip samples (W870924-A)
1,958.97
956.8.050
Acura® electro 956 8-Channel Micropipette, 2.5 – 50 µL, supplied with QC certificate, operating instructions, and Socorex Qualitips® pipette tip samples (W870926)
1,625.64
956.8.050U
Acura® electro 956 8-Channel Micropipette, 2.5 – 50 µL, Initial Package includes electronic pipette, individual QC certificate and operating instructions, pipette charging stand, fast charger with AC adapter and cord, additional battery pack, and Socorex Qualitips® pipette tip samples (W870926-A)
1,900.15
956.8.200
Acura® electro 956 8-Channel Micropipette, 10 – 200 µL, supplied with QC certificate, operating instructions, and Socorex Qualitips® pipette tip samples (W870928)
1,566.82
956.8.200U
Acura® electro 956 8-Channel Micropipette, 10 – 200 µL, Initial Package includes electronic pipette, individual QC certificate and operating instructions, pipette charging stand, fast charger with AC adapter and cord, additional battery pack, and Socorex Qualitips® pipette tip samples (W870928-A)
1,900.15
956.8.350
Acura® electro 956 8-Channel Micropipette, 20 – 350 µL, supplied with QC certificate, operating instructions, and Socorex Qualitips® pipette tip samples (W870930)
1,566.82
956.8.350U
Acura® electro 956 8-Channel Micropipette, 20 – 350 µL, Initial Package includes electronic pipette, individual QC certificate and operating instructions, pipette charging stand, fast charger with AC adapter and cord, additional battery pack, and Socorex Qualitips® pipette tip samples (W870930-A)
1,900.15
956.12.010
Acura® electro 956 12-Channel Micropipette, 0.5 – 10 µL, supplied with QC certificate, operating instructions, and Socorex Qualitips® pipette tip samples (W870932)
1,566.82
956.12.010U
Acura® electro 956 12-Channel Micropipette, 0.5 – 10 µL, Initial Package includes electronic pipette, individual QC certificate and operating instructions, pipette charging stand, fast charger with AC adapter and cord, additional battery pack, and Socorex Qualitips® pipette tip samples (W870932-A)
2,103.35
956.12.050
Acura® electro 956 12-Channel Micropipette, 2.5 – 50 µL, supplied with QC certificate, operating instructions, and Socorex Qualitips® pipette tip samples (W870934)
1,958.97
956.12.050U
Acura® electro 956 12-Channel Micropipette, 2.5 – 50 µL, Initial Package includes electronic pipette, individual QC certificate and operating instructions, pipette charging stand, fast charger with AC adapter and cord, additional battery pack, and Socorex Qualitips® pipette tip samples (W870934-A)
2,032.05
956.12.200
Acura® electro 956 12-Channel Micropipette, 10 – 200 µL, supplied with QC certificate, operating instructions, and Socorex Qualitips® pipette tip samples (W870936)
1,900.15
956.12.200U
Acura® electro 956 12-Channel Micropipette, 10 – 200 µL, Initial Package includes electronic pipette, individual QC certificate and operating instructions, pipette charging stand, fast charger with AC adapter and cord, additional battery pack, and Socorex Qualitips® pipette tip samples (W870936-A)
2,032.05
956.12.350
Acura® electro 956 12-Channel Micropipette, 20 – 350 µL, supplied with QC certificate, operating instructions, and Socorex Qualitips® pipette tip samples (W870938)
1,900.15
956.12.350U
Acura® electro 956 12-Channel Micropipette, 20 – 350 µL, Initial Package includes electronic pipette, individual QC certificate and operating instructions, pipette charging stand, fast charger with AC adapter and cord, additional battery pack, and Socorex Qualitips® pipette tip samples (W870938-A)
Reagent Reservoirs, V-shape, no lid, PP, autoclavable, 70 mL capacity, natural color, 58×132×25 mm, pkg of 20
61.86
330.01L
Reagent Reservoirs, V-shape, hinged lid, PP, autoclavable, 125 mL capacity, natural color, 83×121×34 mm, pkg of 20
106.59
330.08.9
Reagent Reservoirs, 8-channel, no lid, PS, sterile, single wrapped, white, 48 mL capacity, 83×126×13 mm, pkg of 10
116.41
330.12.9
Reagent Reservoirs, 12-channel, no lid, PS, sterile, single wrapped, white, 48 mL capacity, 83×126×13 mm, pkg of 10
116.41
900.901.48U
Power Supply, Fast Charging AC Adapter, 100 - 240V, North American style plug, with cord for connection to charging stands
117.65
320.913.48
Battery Charging Rack, allows the charging and storing of up to three battery packs at a time, for battery packs only (W820019)
309.80
320.903.48
Pipette and Battery Charging Stand, allows the charging and storing of up to three electronic pipettes or battery packs at a time (W820020)
Inquire
900.920.48
Spare Battery Pack, 4.8V NiMH battery pack, blue color, when charged supports at least 3,000 cycles at full plunger travel, fast re-charging (less than 1.5 hours), not subject to a battery memory effect, pkg of 1
276.47
900.922.48
Spare Battery Pack, 4.8V NiMH battery pack, blue color, when charged supports at least 3,000 cycles at full plunger travel, fast re-charging (less than 1.5 hours), not subject to a battery memory effect, (W870001) pkg of 2
276.47
313.02
Pasteur pipette, glass, 2 mL, for 2 mL and 5 mL pipettes, 150 mm length, nonsterile, pkg of 250
Inquire
1.835.631
Adapter nozzle to fit Pasteur pipette for Calibra digital 832, Acura manual 823 and Acura electro 936, 2mL
136.40
1.835.633
Adapter nozzle to fit Pasteur pipette for Acura manual 835 and Acura electro 936, 5mL (W820023)
136.40
322.05
Nozzle filters for pipettes, PP fibers, fits 5 mL Acura manual 835.05, Acura electro 935.05 and 936.05 models, nonsterile, not autoclavable, (W851342) pkg of 250
212.82
322.10
Nozzle filters for pipettes, PP fibers, fits 10 mL Calibra digital 832.10, Acura manual 835.10 and Acura electro 935.10 and 936.10 models, nonsterile, not autoclavable, (851346) pkg of 100